Description
Technical Data LWZ EDI:
( LWZ EDI industrial reverse osmosis water treatment system by reverse osmosis )
- 4-stage reverse osmosis water treatment system
- Output 50 / 100 / 150/ 250 liters/h @ 10°C / varies by customer requirement
- Permeate:
– Conductivity of the product < 0.055 µS/cm > 18.2 MΩ x cm
– Silicate content < 0.1 ppm
– TDS < 1 ppm - Stage 1 reverse osmosis system
- Stage 2 reverse osmosis system
- Stage 3 electrodeionization ( EDI )
- Stage 4 polishing
Stage 1: Reverse Osmosis
- Permeate flow: xxx L/h ( varies by customer requirement ) operating pressure approx. 10 bar
- IEM membrane
- Stainless steel membrane housing
- Stainless steel pump 230V or 400V AC 50/60 Hz
Stage 2: Reverse Osmosis
- Permeate flow: 100 / 150/ 250 /…….. L/h ( varies by customer requirement ) operating pressure approx. x bar
- IEM membrane
- Stainless steel membrane housing
- Stainless steel pump 230V or 400V AC 50/60 Hz
Stage 3: Electrodeionization ( EDI )
To continuously produce ultrapure water, an EDI unit with matching capacity is used.
This technology works via electrodialysis and electrodeionization.
Output: 50 / 100 / 150/ 250
- Conductivity of the product < 0.056 µS/cm > 18.2 MΩ x cm
- Silicate content < 0.1 ppm
- TDS < 1 ppm
Stage 4: Polishing
To guarantee product water quality, Stage 4 is fitted with a so-called polishing cartridge containing lab-grade mixed-bed resin.
This stage serves purely as a quality safeguard with minimal consumption of the mixed-bed resin, since the permeate from the Stage 3 EDI already
has a conductivity of 18.2 MΩ x cm. As a result, mixed-bed resin consumption is significantly lower and lasts much longer compared to conventional systems without EDI and an additional Stage 2 reverse osmosis.
- Mixed-bed resin cartridge
- Diameter 85 mm, height 440 mm
including adapter with 3/8″ quick fitting - Lab-grade resin fill
Main Components LWZ EDI:
- Computer controller type ( IEM PLC control )
- Control cabinet with RS232 or USB interface for programming and remote administration via internet
- Operating hours counter
- Aluminum frame
- Automatic flushing system with solenoid valve
- Solenoid valve on the water inlet at Stage 1
- 2x active digital flow transmitter
(1: product, 1: concentrate ) - 2x active digital pressure transmitter 0-16 bar
( inlet pressure, operating pressure ) - integrated conductivity meter
- Remote monitoring option via internet ( requires a DHCP-capable RJ45 network connection )
Interactive touch screen control system
- Monitoring system with PLC computer control
- 1x transmitter, conductivity display on the permeate
- 3x active digital flow transmitter (1: product, 1: concentrate, 1: recycling)
- 3x active digital pressure transmitter 0-16 bar
Additional features:
- Filter-change indicator
- Status display of pressure and flow
- Quality monitoring
- Fault log
Feed Water Quality:
- Max. salt content in feed water 3000 µS/cm ( design basis 750 µS/cm)
- Min. inlet pressure 1 bar feed water
- Temperature 4° C to 30° C
- Feed water quality: 5µ filtered, drinking water
- Maximum free chlorine: 0.1 mg/l
- Hardness in feed water:
10° dH hardness = 50% recovery
0° dH hardness = 70% recovery




Reviews