3-Stage 250 L/h EDI Ultrapure Water Reverse Osmosis System

Price on request

  • Ultrapure Water Treatment System – Complete Water Treatment Solution for Laboratory Facilities
  • Application area: Research / Laboratory / Medicine
  • Product water quality: 18.2 MΩ x cm ( Class 1 ASTM Standard )
  • Permeate properties:
    • – Conductivity of the product > 18.2 MΩ x cm ( 0.055 µS/cm )
    • – Silicate content < 0.1 ppm- TOC < 30 ppb
  • Basic configuration:
    • Stage 1 reverse osmosis system „free-standing unit Type RO4040x“
    • Stage 2 reverse osmosis system „free-standing unit Type RO4021x“
    • Stage 3 electrodeionization ( EDI )
    • Stage 4 polishing
    • Touch screen control

Description

Specifications 3-Stage 250 L/h EDI Ultrapure Water Reverse Osmosis System:

Type: 3-Stage 250 Liter / Hour Ultrapure Water Reverse Osmosis EDI Water Treatment System
Output: 250 liters / hour (<1 µs/cm @15° C water temperature
Feed water tap water quality:
Softened @ 1 ° dH
Conductivity @ 600µs/cm
Dimensions: approx. 800*1100*1650mm WxDxH ( system only, without options and tanks)
Screen backwash filter
( fine filter ) 1/2″
Dirt particles and foreign matter are reliably filtered out by a fine, dimensionally stable filter screen. Thanks to the backwash function, this filter is completely cleaned again in the shortest possible time.
Pressure reducer Pressure reducer
Backflow preventer The backflow preventer secures the drinking water supply by preventing non-potable water from flowing back into the public water supply network. Installed between the public water supply network and the private consumer network, the backflow preventer creates a separation and thus prevents contact between the two water networks. It protects the public water supply network against downstream overpressure, reverse pressure conditions and water backflow. According to standard EN 12729 per EN 1717.

You do not need this option if you already have a backflow preventer installed.

Softening system Twin softening system Type 200

20°dH 2x 8000 liters,
until the next regeneration,
resin volume per tank 2 x 48.8 liters,
regeneration time 90 min,
salt reserve 100 kg, nominal flow rate per DIN19636 3.6 m³/h,
pressure loss at max. output 0.5 bar,
water pressure min. 1.4 bar, max. 8.6 bar,
water connection 1″ male thread,
power connection V/Hz 230/50/12V, power draw 3 watts, total width with brine tank 710 mm, depth 500 mm, height 1010 mm,
weight without brine tank 90 kg.

 

Pre-filter:
Sediment filter Sediment filter
integrated pre-filter housing 2.5
– incl. filter housing wrench
– 5µ fine filter fitted
– including piping with the system
– with mounting bracket
Type: 3-Stage 250 Liter / Hour Ultrapure Water Reverse Osmosis EDI Water Treatment System (All-in-One solution)
Output: 250 liters / day (< 1 µs/cm @15° C water temperature
without item 2: conductivity < 1 µs/cm @15° C water temperature not accounting for CO2)
Dimensions: 800*1100*1650mm WxDxH. (without further options and tanks)
Aluminum frame with cover on castors
A complete solution from tap water to ultrapure water in a single system with all the required components. Some (optional) components are external — storage tank and options such as softening system, CO2 alkali dosing.
PLC control: IEM computer controller with USB connection and optional remote monitoring and administration via internet
Dry-run protection: Yes
Stage 1 Reverse osmosis water treatment system Type 442
Output: 500 liters / hour (99% rejection @15° C water temperature )
Membrane IEM XLE4040 ( extra low energy PA membrane )
Membrane housing: Stainless steel pressure vessel, PE end caps
Pump: Stainless steel centrifugal pump 400V AC, low-noise, multi-stage
Mounting – Design: mounted on aluminum frame
Flow sensors: 2x active digital flow transmitter
(1: product, 1: concentrate)
Pressure sensors: 2x active analog pressure transmitter 0-16 bar
( inlet pressure, operating pressure )
Dry-run protection: Yes
Flushing system: Automatic flushing system ( solenoid valve )
Conductivity display: conductivity monitoring on the permeate
Permeate rejection with permeate reject system
Stage 2 Reverse osmosis water treatment system Type 441
Output: 300 liters / hour (99% rejection @15° C water temperature )
Membrane IEM XLE4040 ( extra low energy PA membrane )
Membrane housing: Stainless steel pressure vessel, PE end caps
Pump: Booster pump, stainless steel P211 / 640 W
Conductivity display: conductivity display on the permeate
Mounting – Design: Aluminum frame
Flow sensors: 1x active digital flow transmitter
(1: product )
Pressure sensors: 1x active analog pressure transmitter 0-16 bar
( operating pressure )
Dry-run protection: Yes
Stage 3 EDI ( internal)
Output: 250 liters / hour <1 µs/cm
Type: inline electrodeionization system
Output 250 liters / hour, 5 – 40° C feed water temperature
Type: industrial
EDI module – Industrial
230 V AC 50 Hz
EDI conductivity monitoring
Polishing filter DI (polishing cartridge) external material GFK
1x 350 liters / hour <1 µs/cm
Resin fill ( mixed-bed anion + cation ) laboratory grade
1x fill
Conductivity display on the permeate as text output
including adapter
1x fill
Conductivity display on the permeate as text output

 

including adapter
Feed water inlet: Solenoid valve on the water inlet 230 V 50Hz
Connections: Permeate 3/4″ male thread / Concentrate 3/4″ male thread / Feed water 3/4″ male thread
Shutdown: Float switch
Limits: Max. salt content in feed water 3000 µS/cm ( design basis 750 µs/cm)
Min. inlet pressure 1 bar feed water
Temperature 4° C to 30° C
Feed water quality: 5µ filtered, drinking water
Maximum free chlorine: 0.1 mg/l
Hardness in feed water:
10° dH hardness = 50% recovery
0° dH hardness = 70% recovery
Standby flushing To prevent an increase in TOC (Total Organic Carbon) content in standby mode, the system must be operated at least once every 12 hours.
If the system remains in standby mode for more than 12 hours, it switches to production mode and the product water is directed to drain via a motorized 3-way stainless steel valve.
This function is programmed into the PLC.
Conveying system with storage tank:
Storage tank: Product water storage tank 5000 L ( external)
(storage tank made of high-quality PE)
Max. distance from RO system 5 m
Diameter 2030 mm
Height 1800 mm
Top connections 2″ and 27 mm
Bottom connections 2″ and 3/4 mm. Note: ( only the tank is supplied; all piping from the system to the tank and the conveying pump must be carried out on site by the operator of the system )
CO2 trap CO2 trap
serves to prevent the ultrapure water ( in the product water tank ) from absorbing CO2 through ventilation during withdrawal; the air is routed through the CO2 trap.
2.5×10 inch laboratory grade
Level measurement Level sensor transmitter ( transducer )
Installed and programmed in the PLC control
precise level measurement via a pressure-based method. Tank full ( stop production)
Tank half-empty ( start production)
Tank empty ( alarm )
programmed in the system’s PLC. Technical parameters:
Input voltage : DC 24V DC
Current draw : 200 uA
Output : 1-10 V
Response time : 500 ms
Operating temperature : 0 ~ 45 degrees
Pump: Booster pump
Conveying pump Type: IEMHJ550Q/max.: 58 L/min ( 3.5 m³/h )
H/max.: 45m ( 4.5 bar)
Motor 220 – 240 V~/50 Hz. Item description
HJ550 stainless steel pump, material stainless steel 1.4301 ( AISI 304)
Suction lift approx. 9 m
Delivery head approx. 45 m
Capacity: 3500 l/h
Max. delivery pressure approx. 4.5 bar
Rated voltage: 220 – 240 V~/50 Hz 550 W ( 1 Hp)
integrated thermal overload protection
Maximum water temperature: 60°C
Suction connection: 1″ female thread
Pressure connection: 1″ female thread
Insulation class : B
IP 44
Max. ambient temp. +40°C
All figures @ 2900 rpm ( 50 Hz). Mounted on aluminum frame
controlled by the reverse osmosis system’s own PLC control
Pressure tank Pressure tank
1 x 20 liter stainless steel pressure tank for compensation and as storage tank
20 liter total volume of air and water; water volume depends on the inlet pressure ( 10 liters water @ 5bar )
Including fittings, valves and hoses
Direct withdrawal Direct withdrawal connection on the unit
Mounting – Design: mounted on aluminum frame
UV disinfection Ultraviolet disinfection 50 L/min
Ultraviolet system UV system with failure monitoring ( built-in )
Stainless steel water sterilizer ( UV system) 50 watt
UV water sterilizer – flow rate: 50 liters per minute
Quartz lamp 220 V, 50Hz, 50 watt –
Fault signal via PLC with display in the control cabinet
on failure
Circulation pump Conveying pump Type: IEMHJ550

Q/max.: 58 L/min ( 3.5 m³/h )
H/max.: 45m ( 4.5 bar)
Motor 220 – 240 V~/50 Hz

Item description
HJ550 stainless steel pump, material stainless steel 1.4301 ( AISI 304)
Suction lift approx. 9 m
Delivery head approx. 45 m
Capacity: 3500 l/h
Max. delivery pressure approx. 4.5 bar
Rated voltage: 220 – 240 V~/50 Hz 550 W ( 1 Hp)
integrated thermal overload protection
Maximum water temperature: 60°C
Suction connection: 1″ female thread
Pressure connection: 1″ female thread
Insulation class : B
IP 44
Max. ambient temp. +40°C
All figures @ 2900 rpm ( 50 Hz)

Mounted on aluminum frame
controlled by the reverse osmosis system’s own PLC control

Filter 0.2µ Sterile filter 0.2µ
Filter type 0.2 µ cartridge filter
Material: Polypropylene
2.5 inch ( filter area 0.65 m² )
Operating temp. up to 65° C
Operating pressure 4.2 bar ( max. differential 2 bar)
with 2 differential pressure gauges
Monitoring interactive monitoring system and remote administration
Touch Screen interactive touch screen system + remote monitoring
Monitoring system with PLC computer control
control cabinet with RS232 or USB interface for programming and remote administration via internet
1x conductivity display transmitter on the permeate
3x active digital flow transmitter
(1: product, 1: concentrate, 1: recirculation)
3x active digital pressure transmitter 0-16 bar. Further functions:
– status display of pressure and flow
– quality monitoring
– error log. The current status of the system can be monitored remotely by the customer as needed with the help of client software ( monitoring ). A DHCP-capable internet connection is required at the system via a connected RJ-45 cable on the customer’s side for this.
Changes to the software can only be made remotely by the manufacturer ( IEM ) after mutual agreement. The customer does not receive access to the PLC programming; this remains reserved for the manufacturer ( IEM ).

Other Related Products:

2-Stage Reverse Osmosis Water Treatment System RO806-804

LabWasserZentrale Ultrapure Water Treatment System

2-Stage Reverse Osmosis Water Treatment System RO Free-Standing Unit

3-Stage 225 L/h EDI Ultrapure Water Reverse Osmosis System

3-Stage 20 L/h EDI Ultrapure Water Reverse Osmosis System

Additional information

System Type
Gehäuseart
Housing Type

Output

Reviews

There are no reviews yet.

Be the first to review “3-Stage 250 L/h EDI Ultrapure Water Reverse Osmosis System”

Your email address will not be published. Required fields are marked *